Measurement Specialties Inc. Model 9116 User’s Manual
Page 79 www.MEAS-SPEC.com
Example:
Send TCP/IP command to module (via its connected socket) requesting the most
recent calibration adjustment’s offset and gain terms (cc=00-01), and the
adjacent factory-determined transducer coefficients C0 through C4 (cc=02-06) for
transducer 1: Data requested in ASCII-hex format representing the internal
binary floating point format.
u10100-06
Response returned is:
3B200A6E . . 00000000
Note
The maximum response size is 300 characters. If the requested
range of coefficients requested exceeds this, the module will return
an “N07" error response.